Tuesday, February 19, 2019

Medtronic, Inc. Recalls Dual Chamber Implantable Pulse Generators (IPGs)

Medtronic, Inc. Recalls Dual Chamber Implantable Pulse Generators (IPGs) Due to Possible Circuit Error

 

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death

Recalled Product(s):

  • Medtronic, Inc. Dual Chamber Implantable Pulse Generators (IPGs)
  • Model Names:
    • Adapta, Versa, Sensia, Relia, Attesta, Sphera, and Vitatron A, E, G, Q series
  • Manufacturing Dates: March 2, 2017, to December 18, 2018
  • Distribution Dates: March 6, 2017, to January 7, 2019
  • Devices Recalled in the U.S.: 13,440

Device Use

Medtronic’s Dual Chamber Implantable Pulse Generators (IPGs) are implanted cardiac pacemakers used to provide stimulation to increase heart rate in patients with a slow heart rhythm (bradycardia) or no heart rhythm. The pulse generator is the small implanted unit containing the battery and other electronic parts. The pulse generator must be used with insulated electrode wires called leads. These devices are designed to be used in addition to routine clinical monitoring by a health care professional.

Reason for Recall

Medtronic is recalling its dual chamber IPGs due to the possibility of a software error that can result in a lack of pacing. Patients and physicians cannot predict whether and when this software error might occur. A lack of pacing could result in patients experiencing slow heart beating, low blood pressure, and symptoms such as light headedness, fainting, and even death.

Who May be Affected

  • Hospitals and health care professionals using Medtronic’s Dual Chamber Implantable Pulse Generators to provide pacing support to treat patients with bradycardia.
  • Patients and caregivers of patients receiving pacing support using a Medtronic Dual Chamber Implantable Pulse Generator.

What to Do

Beginning January 17, 2019, Medtronic’s Field Representatives have hand-delivered “Field Corrective Action Notification” letters to implanting and follow-up physicians. The notification provided the following instructions:
Health Care Providers
  • Medtronic recommends programming to a non-susceptible pacing mode as the primary mitigation for patients implanted with an affected device until the software update has been installed.
  • Additional patient risk assessment and programming recommendations are included in Medtronic’s advisory letterdisclaimer icon.
  • In addition to the field correction notification, Medtronic’s Field Representatives will:
    • Hand deliver an additional “Supplemental Letter” alongside an “Urgent Medical Device Recall” letter to physicians with patients whose devices have shown evidence of a pacing pause that may be related to this circuit error.
    • Request physicians return all unused and unopened affected product to Medtronic for replacement.
    • Confirm notifications are received by implanting or follow-up physicians using Medtronic’s electronic tracking system in conjunction with confirmation via a paper form.
    • Conduct effectiveness checks to ensure that all identified implanting or follow-up physicians and risk managers have been notified or proof of at least three attempts to notify them is obtained.
Patients
  • Patients remaining in a susceptible mode should seek immediate medical attention if experiencing any new or unexpected symptoms consistent with a pause in pacing.

Contact Information

Customers who have questions or need additional information regarding this recall may contact Medtronic’s Technical Services at 1-800-505-4636.

Date Recall Initiated

January 17, 2019

Additional Resources:

 

No comments:

GA

Make sure you know about all recalls & keep your family safe, Subscribe Free via email

Enter your email address Or Click on FeedBurner Icon (Right) to subscribe via Your News Reader. Your Email is never shared:

Delivered by FeedBurner