Class I Recall - Smiths Medical AKA Medex Cardio-Pulmonary Recalls Sterile Saline and Sterile Water for Inhalation Due to Potential Exposure to Infectious Agents

Medex Cardio-Pulmonary, doing business as Smiths Medical, Recalls Sterile Saline and Sterile Water for Inhalation Due to Potential Exposure to Infectious Agents as a Result of Leaking Containers.

The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death
Prefilled Humidifier with Adapter Portex 350 mL 0352 Each/1
One of the recalled products Portex 350 ml Sterile Water Humidifier w/5psi Adapt

Recalled Product(s):

  • Select serial and lot numbers of Smiths Medical Sterile Saline and Water (listed below)
  • Manufacturing Dates: August 30, 2013, to February 10, 2017
  • Distribution Dates: August 30, 2013, to February 10, 2017
  • Full list of affected devices
  • Devices Recalled in the U.S.: 939,364

Device Use

The sterile saline and water are respiratory humidification products used in patients undergoing respiratory therapy, tracheal wash, or wound cleaning.

Reason for Recall

Smiths Medical is recalling the sterile saline and sterile water products for inhalation due to the potential exposure to infectious agents (bacillus infantis and staphylococcus epidermidis) because of damage to the containers used to package the finished products. Use of these products in patients could result in infection and may require treatment with antibiotics. Serious or untreated infections could result in patient death.
The company initiated a voluntary recall on September 5, 2017. That recall covered several products, including some that are outside the scope of this notice. The FDA is auditing the recall to ensure the company has notified all affected customers and that affected product has been returned. Based on the available information, the FDA is now classifying the action regarding the affected products (listed below) as a Class I recall.

Who May be Affected

  • Hospitals and health care professionals using Smiths Medical Sterile Saline or Water distributed from August 30, 2013, to February 10, 2017.
  • Patients receiving respiratory therapy or tracheal lavage with Smiths Medical Sterile Saline or Water distributed from August 30, 2013, to February 10, 2017.

What to Do

On September 5, 2017, Smiths Medical sent affected customers an "Urgent Medical Device Recall Notice" informing them of affected sterile water and saline product models and directing them to locate, determine the affected sterile saline or water products in their possession, and to complete the “Urgent Medical Device Recall Response Form” attached to the recall notice. Customers were also instructed to return all affected product to Stericycle for processing.
On January 7, 2019, distributer Medline Industries Inc. initiated a sub-recall for privately labeled product models and lots, affected by the September 5, 2017, Smiths Medical recall notice, that were not expired and which may be potentially contaminated with the infectious agents.
The notice asked customers to:
  1. Locate the affected Sterile Water, Saline and Eye Wash in their possession by referring to the specific model and lot numbers listed in the Urgent Medical Device Field Recall Response Form.
  2. Determine the number of affected devices in their possession and complete the Urgent Medical Device Field Recall Response Form attached to this letter within 10 days of receipt and send it to smithsmedicalste00100@stericycle.com. The form must be returned even if customers do not have any of the affected Sterile Water, Saline and Eye Wash in their possession. Product credit will be processed once the Urgent Medical Device Field Recall Response Form and affected product is received. All affected devices must be returned to Stericycle for processing. Pre‐paid shipping labels were included with the Urgent Medical Device Recall Notice.
  3. Package the affected devices and include a copy of the completed Urgent Medical Device Field Recall Response Form inside EACH BOX of the returned devices to obtain credit for the returned devices. Make sure boxes are sealed and labeled with the facility name prior to shipping devices to Stericycle.
Distributors are asked to immediately notify customers who received the affected product of this Recall.

Contact Information

Customers with questions regarding this notification may contact Stericycle via email at smithsmedicalste00100@stericycle.com.

Date Recall Initiated

September 5, 2017

Full List of Affected Devices

  • Portex 350 ml Sterile Water Humidifier w/5psi Adapt (Model # 352 Lot Numbers: A214, A215, Z263, Z532, and Z555)
  • Portex 550 ml Sterile Water Humidifier w/5psi Adapt (Model # 552 Lot Numbers: A054, A089, A090, and Z262)
  • Portex 1000 ml Sterile Water USP Pour Bottle (Model # 1065 Lot Numbers: B209 and Z225)
  • Portex 500 ml Sterile Water USP Pour Bottle (Model # 1565 Lot Number: Y371)
  • Intermed 350 ml Sterile Water Humidifier w/5psi Adapt. (Model # 0352IMJ Lot Numbers: A457, A597, B157, B236, Z589, Z655, Z656, and Z661)
  • Portex 550 ml Sterile Water Humidifier w/5psi Adapt. (Model # 0552C Lot Number: Z370)
  • Intermed 550 ml Sterile Water Humidifier w/5psi Adapt. (Model # 0552IMJ Lot Numbers: Z588 and Z597)
  • Medline 300-350 ml Sterile Water Humidifier w/5psi Adapt. (Model # HCS00300 Lot Number: Y576)
  • Medline Prefilled 350 ml Sterile Water (Model # HCS00350 Lot Numbers: A055, A056, A057, A058, A103, A176, A455, B530, B531, B532, B533, Z101, Z534, Z553, and Z554)
  • Portex Unit Dose 5 ml Normal Saline (0.9%) (Model # R0059 Lot Number: B360), K820227
  • Portex Unit Dose 15 ml Normal Saline (Model # R0159 Lot Numbers: A661, B067, A526, A536, A569, and B201)
  • Portex 5ml Normal Saline (0.9%) Unit Dose (Model # UD9005 Lot Number: B515), K820227
  • Medline Prefilled 550 ml Sterile Water (Model # HCS00550 Lot Numbers: A092, B205, B534, and Z205)

Thursday, August 6, 2009

Beef Packers, Inc., a Fresno, California, Recalls 825,769 pounds of ground beef due to salmonellosis Outbreak

Recall Release CLASS I RECALL
FSIS-RC-041-2009 HEALTH RISK: HIGH

Congressional and Public Affairs
(202) 720-9113
Brian K. Mabry

WASHINGTON, August 6, 2009 - Beef Packers, Inc., a Fresno, Calif., establishment, is recalling approximately 825,769 pounds of ground beef products that may be linked to an outbreak of salmonellosis, the U.S. Department of Agriculture's Food Safety and Inspection Service (FSIS) announced today.

The products subject to recall include:
  • 60 pound cases of "GRD Beef Sirl Fine 90/10." Each case bears the identifying case code W69012 with Use/Freeze by dates of 06/23/09, 06/26/09, 06/28/09, 06/29/09, 06/30/09, 07/03/09, 07/05/09, 07/06/09, 07/07/09, 07/10/09 and 07/11/09.
  • 60 pound cases of "Grnd Bf Fine Sir 90/10 10/60 H". Each case bears the identifying case code W69064 with Use/Freeze by dates of 06/26/09, 07/03/09, 07/06/09, 07/07/09 and 07/14/09.
  • 60 pound cases of "Grnd Beef 90/10 Fine 60." Each case bears the identifying case code W69063 with Use/Freeze by dates of 06/23/09, 06/26/09, 06/30/09, 07/03/09, 07/07/09, 07/10/09 and 07/11/09.
  • 80 pound cases of "Grnd Beef Fine 91/09 10#/80." Each case bears the identifying case code W69108 with Use/Freeze by dates of 06/30/09, 07/06/09, 07/08/09 and 07/13/09 .
  • 60 pound cases of "Grnd Beef Fine 93/07 10/60." Each case bears the identifying case code W69363 with Use/Freeze by dates of 06/23/09, 06/26/09, 06/28/09, 06/29/09, 06/30/09, 07/03/09, 07/07/09, 07/10/09 and 07/11/09.
  • 60 pound cases of "Grnd Beef Fine 93/07 10#/60 H." Each case bears the identifying case code W69360 with Use/Freeze by date of 06/26/09, 07/03/09, 07/06/09, 07/10/09 and 07/14/09.
  • 40 pound cases of "GRD BEEF FINE 96/04-10#/40." Each case bears the identifying case code W69602 with Use/Freeze by dates of 06/23/09, 06/30/09, 07/03/09, 07/07/09, 07/10/09 and 07/11/09.
The ground beef products were produced on various dates ranging from June 5, 2009 through June 23, 2009 and bear the establishment number "EST. 31913" printed on the case code labels. The ground beef products were distributed to retail distribution centers in Arizona, California, Colorado and Utah. Because these products were repackaged into consumer-size packages and sold under different retail brand names, consumers should check with their local retailer to determine whether they may have purchased any of the products subject to recall.

As a result of an ongoing investigation into an outbreak of Salmonella Newport associated with ground beef products, the Colorado Department of Public Health and Environment (CDPHE) notified FSIS of the situation. Epidemiological and traceback investigations conducted by FSIS and CDPHE determined that there is an association between the fresh ground beef products and illnesses reported in Colorado. The illnesses were also linked through the epidemiological investigation by their uncommon pulsed-field gel electrophoresis (PFGE) pattern found in PulseNet, a national network of public health and food regulatory agency laboratories coordinated by the Centers for Disease Control and Prevention.

FSIS would like to remind consumers of the importance of following food safety guidelines when handling and preparing raw meat. Ground beef should be cooked to a safe minimum internal temperature of 160° Fahrenheit.

This particular strain of Salmonella Newport is resistant to many commonly prescribed drugs, which can increase the risk of hospitalization or possible treatment failure in infected individuals.

Consumers with questions about the recall should contact the company's Consumer Line at (877) 872-3635. Media with questions should contact company Director of Communications Rebecca Hayne (316) 291-2126.

Consumption of food contaminated with Salmonella can cause salmonellosis, one of the most common bacterial foodborne illnesses. Salmonella infections can be life-threatening, especially to those with weak immune systems, such as infants, the elderly, and persons with HIV infection or undergoing chemotherapy. The most common manifestations of salmonellosis are diarrhea, abdominal cramps, and fever within eight to 72 hours. Additional symptoms may be chills, headache, nausea and vomiting that can last up to seven days.

Consumers with food safety questions can "Ask Karen," the FSIS virtual representative available 24 hours a day at AskKaren.gov. The toll-free USDA Meat and Poultry Hotline 1-888-MPHotline (1-888-674-6854) is available in English and Spanish and can be reached from l0 a.m. to 4 p.m. (Eastern Time) Monday through Friday. Recorded food safety messages are available 24 hours a day.

GA

Make sure you know about all recalls & keep your family safe, Subscribe Free via email

Enter your email address Or Click on FeedBurner Icon (Right) to subscribe via Your News Reader. Your Email is never shared:

Delivered by FeedBurner