Cook Medical Inc. Recalls Transseptal Needle Due to Risk of Detached Plastic Fragments
The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.
- Recalled Product(s): Transseptal needle
- Lot Number: 8833687
- Model: TSNC-18-71.0
- Manufacturing Date: April 23, 2018
- Distribution Dates: May 30, 2018 to November 5, 2018
- Devices Recalled in the U.S.: 97
Device Use
The Transseptal Needle is used by surgeons to access the left side of a patient’s heart during cardiac procedures. Fluoroscopy medical imaging and a guide catheter help the surgeon place the needle in the heart.
Reason for Recall
Cook Medical is recalling one lot of the Transseptal Needle due to a manufacturing error that resulted in some needle tips missing the back bevel that creates a sharp tip. Without a back bevel, the needle tip could damage the inside of the introducer sheath during insertion of the needle resulting in detached plastic fragments. These fragments could potentially enter the patient’s bloodstream and result in serious adverse health consequences such as a longer procedure to retrieve the plastic pieces, injury to blood vessel walls, embolism, stroke, or death.
Who is affected?
- Health care providers using this needle during cardiac procedures
- Patients undergoing cardiac procedures with the use of these needles
What to Do
On February 1, 2019, Cook Inc. sent an Urgent Medical Device Recall notification letter to affected customers. The notice asked customers to:
- Share notice with appropriate staff
- Identify and remove any affected transseptal needles and quarantine affected needles
- Immediately stop all distribution and use of affected products
- Return any affected product(s) to Cook Inc. Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit
- Complete and return the Acknowledgement and Receipt Form by fax at 812.339.7316 or email at FieldActionsNA@CookMedical.com
Contact Information
Health care professionals and distributors with questions are instructed to contact Cook Medical Customer Relations by phone at 800-457-4500 or 812-339-2235, Monday through Friday between 7:30 A.M. and 5:00 P.M. (Eastern Time) or by email to CustomerRelationsNA@CookMedical.com.
Date Recall Initiated
February 1, 2019